Set of instruments for test protection against accidental touch
The set of instruments is designed for testing electrical products to protect it from:
- electrical shock under normal operating conditions;
- solid objects penetration;
- human access to dangerous parts of the products and
- contact with foreign objects into inside of the products.
The set of instruments meets the requirements according to the standards:
- HOST 14254-96 (MEK 529-89). Degrees of protection provided by protective shells (IP code).
- HOST 25861-83 (ST REV 3743-82). Computing machines and data processing systems. Electrical and mechanical safety requirements and testing procedures.
- DSTU IEC 60335-1:2004 Household and similar electrical appliances. Safety. Part 1. General requirements.
- DSTU IEC 60065:2009 Audio, video and similar electrical appliances. Safety requirements.
The set of instruments consists of the following components:
- set of testing instruments – 10 units;
- signaling unit;
- case for storing testing instruments and signaling unit.
Nomenclature and characteristics of the testing instruments
Testing instrument | According to the standard | Testing efforts, H |
1. Tough test finger | DSTU IES 60065 | 50- |
2. Swivel test finger | HOST 14254 HOST 25861 DSTU IES 60065 DSTU IES 60335-1 |
-
10 - 20 |
3. Probe access | HOST 14254 | 50 |
4. Test rod 2.5 mm diameter | HOST 14254 | 3 |
5. Test rod 2.5 mm diameter 1,0 mm | HOST 14254 | 1 |
Test pin | DSTU IES 60065 | 10 |
Test nail | DSTU IES 60335-1 | 10 |
Tough finger | DSTU IES 60335-1 | 50 |
Test probe | DSTU IES 60335-1 HOST 25861 |
- |
Test rod | DSTU IES 60335-1 HOST 25861 |
- |
NOTE Deviation values of testing efforts within ± 10% for all instruments.
Technical characteristics of the signaling unit
Signaling unit power supply:
Voltage | 220 V ± 22 V |
Frequency | 50 Hz ±1 Hz |
The value of the output test voltage of DC generated by signaling unit | 40 V ± 2 V |
Power consumption, no more than | 6 W |
Overall dimensions of the case (set of instruments during storage), no more than:
|
460 mm
360 mm
165 mm
|
Weight of case with instruments and signaling unit, no more than | 7,0 kg |
Vitalii KELVYCH
Head of the department of experimental production and measurement instruments repair
Tel./Fax: +380 44 526-75-46; +380 50 443-62-31
e-mail: This email address is being protected from spambots. You need JavaScript enabled to view it.