Set of KSS-02 neutral light filters
It is designed for verification of photometers, biochemical analyzers, photometers of general purpose, spectrophotometers of the ultraviolet, visible and near-infrared regions of the spectrum.
   
Metrological characteristics:
| 1. Nominal value of the directional transmission coefficient (T) | 5, 10, 15, 30, 70 and 90% | 
| 2. Extended uncertainty | ±(0,3-0,5)% | 
| 3. Working spectral range | from 200 to 2500 nm | 
| 4. Extended uncertainty of measure PS-7 absorption bands wavelengths | ± 0,3 nm | 
Delivery set:
- six neutral light filters;
- ZhS-4 light filter;
- measure PS-7;
- certificate of calibration.
Consumers:
Territorial bodies of the Ministry of Economy of Ukraine, metrological services of enterprises and organizations.
CONTACT FOR PRODUCTION:
Vitalii KELVYCH
Head of the department of experimental production and measurement instruments repair
Tel./Fax: +380 44 526-75-46; +380 50 443-62-31
e-mail: 
CONTACT FOR CALIBRATION:
Head of the department of research and production departmnet of optico-physical quantities
Tel./Fax: (044) 526-36-98
E-mail: 